Heat Load — wiring heat budget
How much heat the cables bring to each stage of a dilution refrigerator. For each section between two stages the calculator integrates the material’s thermal conductivity k(T) over the cross-section and divides by the length, as in Eq. 1 of Krinner et al. (2019), and sums over all lines.
Stages
| stage | T (K) | cable from the stage above (mm) | available cooling power (W) |
|---|---|---|---|
| RT | |||
| 50K | |||
| 4K | |||
| Still | |||
| CP | |||
| MXC |
Defaults: Bluefors XLD400 as reported by Krinner et al., Table 1 (power at the temperature given in the source). Replace them with your fridge’s values.
Lines
Load per stage
† uses k(T) below the range of the NIST fits (linear extrapolation in T): an order-of-magnitude estimate, about 2× lower than Krinner’s.
Check: for a UT-085 SS-SS coax with attenuators at 4K, CP and MXC the calculator gives 23.5–27.2 mW on the 50 K stage and 0.44–1.92 mW on the 4 K stage; Krinner et al. predict 24–27 mW and 0.4–1.7 mW and measure 45(23) mW and 1.0(2) mW (Table 2).
Not modelled: radiation, active load (signals and currents), contact resistances, heat from components, NbTi and CuNi conductivity (not in the NIST database).
NIST k(T): trc.nist.gov/cryogenics · S. Krinner et al., EPJ Quantum Technol. 6, 2 (2019), doi:10.1140/epjqt/s40507-019-0072-0 · UT-085C: Micro-Coax datasheet.